Assays Make Opiate Analysis Pain-Free

High-throughput extraction of opiates from urine using Strata-X-C SPE and analysis by GC-MS or LC-MS-MS reduces sample processing time and improves results.

by Michael Rummel, Matthew Trass, Michael Campognone, and Sky Countryman, Phenomenex, Inc.

Click to Enlarge.
Figure 1. Optimization of wash and elution conditions for SPE. Click to Enlarge.
The confirmation of abused opiates by GC-MS or LC-MS-MS can be complicated by the presence of endogenous compounds and non-target opiates with similar mass ions. This technical note describes a new solid phase extraction (SPE) method for the extraction and analysis of opiates that dramatically reduces overall sample processing time and improves chromatographic results by either GC-MS or LC-MS-MS.

The prescription of opiate drugs is common for the treatment of chronic pain. Due to the addictive nature of these drugs, many opiates are regulated under Schedule I and routinely tested as part of pre-employment screening or workplace performance monitoring in the United States. The confirmation of these analytes by GC-MS can be complicated by two factors: 1) removal of endogenous compounds from urine that can co-elute with target analytes, and 2) the presence of non-target opiates that have similar mass ions. The objective of this work was to develop a high-throughput assay for opiates in urine utilizing solid phase extraction (SPE) for sample clean up and analysis by either GC-MS or LC-MS-MS.

Method 

Click to Enlarge.
Table 1. SPE Protocol. Click to Enlarge.
Urine samples were spiked at 2,000 ng/mL with morphine, codeine, and other commonly encountered opiate related interferences. A 2-mL aliquot of urine was then spiked with internal standards and 400-µL concentrated HCI. Hydrolysis was performed by heating the sample at 90 C for 40 min in a heating block, or in an autoclave for 15 min on a liquid cycle. After samples were cooled, they were centrifuged for 10 min at 2000 rpm, and the pellet was discarded. The hydrolyzed samples were then buffered and extracted using a 30-mg/3-mL SPE tube containing a polymeric cationic exchange sorbent under the conditions specified in Table 1. Wash conditions were optimized to remove polar cationic contaminants that interfered with the GC-MS analysis. Separation by GC-MS of morphine and codeine from other opiate related inferences was evaluated using two different derivatization procedures. Samples were simultaneously analyzed by LC-MS-MS on an Applied Biosystems API 3000 without the need for derivatization.


Results and discussion

Click to Enlarge.
Figure 2. GC-MS analysis of samples derivatized using BSTFA. Click to Enlarge.
Due to the increase in sensitivity provided by newer MS instrumentation, especially LC-MS-MS, the volume of urine needed for extraction in order to meet SAMHSA detection limits can be reduced. We were able to reduce the volume to 1 mL of urine, which was then diluted 1:1 with a buffer solution to a total volume of 2 mL. This volume allowed us to use a new optimized SPE format with 30 mg of polymeric cation exchange sorbent packed into a 3-mL tube. Polymeric sorbents offer several important advantages over silica-based sorbents, including improvements in tube to tube reproducibility due to metal content in silica, the reduced susceptibility to de-conditioning if the sorbent bed dries out between processing steps, and the improved flow characteristics.

By using a 30-mg bed mass, we were significantly able to reduce the solvent volume required for conditioning, wash, and elution steps. This reduced solvent volume saves both time and money significantly for subsequent blow down steps prior to derivatization. The recoveries for morphine and codeine were 97% and 98% respectively, with % CVs <4% for replicate injections over three different concentrations, indicating that the sorbent mass was sufficient to reproducibly perform the extraction.

Click to Enlarge.
Figure 3. GC-MS analysis of samples derivatized using propionic acid anhydride.  Click to Enlarge.
When developing the method, several different wash and elution options were explored that provided visually different results. The opiate species are cationic and were retained via ion-exchange mechanisms on the Strata-X-C sorbent. Washing with acidified water solutions helps to remove polar acidic and neutral components. In this case, the use of buffer significantly reduced the presence of a yellowish impurity (Figure 1). However, even the use of a buffer solution was not sufficient to completely remove this impurity. Additional experiments suggested that this impurity was insoluble in strong organic solutions. By changing the elution solvent system from basified methanol to a mixture of 7:2:1 ethyl acetate / isopropanol / ammonium hydroxide, we were able to almost completely remove this impurity from the final extracts.

GC-MS remains the golden standard for SAMHSA opiate testing. The large number of routinely prescribed opiate related compounds that cause positive hits in the ELISA screening assay requires a GC-MS methodology that is able to accurately confirm the presence of morphine or codeine. To improve this separation, the use of a newly developed GC stationary phase specifically designed for the analysis of drugs of abuse was evaluated. To further improve resolution, we investigated the difference in chromatographic performance between BSTFA and propionic acid anhydride derivatized samples.

Click to Enlarge.
Table 2. Mass Ions for Chromatographic. Click to Enlarge.
The elution order of morphine and codeine was reversed on the ZB-Drug-1 vs. standard 5 and 5-ms phases. BSTFA derivatives were well resolved from all common co-eluting impurities, though problems were observed with norcodeine (Figure 2). The use of BSTFA also required the formation of an oxime derivative to prevent multiple reaction products. By forming a propionic acid anhydride derivative, resolution of morphine and codeine from common interferences was greatly improved (Figure 3). Furthermore, we found that we were able to eliminate the oximation step, further reducing processing time.

The SPE protocol presented here do not require derivatization and can also be used to prepare samples for LC-MS-MS (Figure 4 and Table 2). LC-MS-MS has been reported to be at least one order of magnitude more sensitive than GC-MS for drugs of abuse. This would suggest that the volume of urine needed for SPE could be further reduced. If volumes of diluted urine could be reduced to about 1.5 mL, it would allow for the use of a 96-well plate SPE format, which can be easily adapted for automation. Automation could be used to streamline workflows and reduce error in sample processing.

Conclusion

A new procedure for the extraction and analysis of opiates from urine was developed that dramatically reduced overall sample processing time and improved chromatographic results by either GC-MS or LC-MS-MS.

For more information, contact Sky Countryman, Manager, Product Management and Applied Technology, Phenomenex, Inc., at SkyC@phenomenex.com or 209-267-1016


Contact Editor Print Article Email To Friend

Comments
Type Your Comment...
Name:   
E-mail (optional) *E-mail addresses do not publish to the site.
Comment:   


Chrom Notebook

michelle
Michelle Longo
Asst.
Managing Editor

Slow and Steady Wins the Race
In the 100 years since Marie Curie won the Nobel Prize in Chemistry, women have made leaps and bounds in science and engineering. Following the extreme example of Madame Curie, women have exponentially raised their position in S&E in the last few decades. In a field previously (and traditionally) dominated by men, women seem to be slowly but surely taking over—or at least leveling the playing field. ...continue



Application Notes
  • Heparin Characterization
  • Sensitive, Ultrafast HPLC Fluorescence Detectors
  • Water’s Impact on Ion Chromatography
More Application Notes

Most Viewed Content

Advantage Business Media © Copyright 2011 Advantage Business Media
Privacy Policy |  Terms & Conditions |  Advertise with Us