Decontamination Concepts for CO2
A User’s Perspective by Daniela Maurer, Dipl. Biochem., Scientific Product Management, BINDER GmbH, Tuttlingen
Microbial contamination—caused by bacteria, bacterial spores, viruses, mycetozoa, yeast or other microorganisms—frequently presents a major risk in cell culture experiments. Because this contamination does not necessarily occur together with the overgrowth of the cultivated cell type, it is often detected too late. More subtle effects, such as the deprivation of essential nutrients and segregation of microbial metabolites, are caused by slight changes in the pH, which for human and mammalian cells has to be kept within a range of 7.4 to 7.6; the resulting hyperacidity of the culture medium then slows down the growth rate and decreases confluence. Changes in host cell morphology and even genetic changes such as chromosomal aberration and translocation can, for instance, be caused by mycoplasma infection. In extreme cases, a single germ can destroy weeks or months of intricate research work efforts. Causes for the introduction of germs or dispersion of contamination can be innumerable: use of cell lines, media, serum or other reagents with hidden contamination, airborne bacteria, or laboratory equipment that was not properly disinfected or sterilized, or contamination that is accidentally introduced by lab technicians.
Because checking for the presence of germs frequently involves complicated and tedious procedures, measures for contamination control must be initiated.
The Importance of Contamination Control When Working with Cell Lines and Primary Cultures
Binder's C 150 CO2 |
In view of the significant progress in the area of sensitive cell culture applications, such as tissue engineering or regenerative cell and tissue therapy, the requirements for CO2 incubators have changed.
Highest standards are thus applied to the perfection and reliability of the entire process chain, in which the CO2 incubator occupies a key position, because it must replicate the natural in vivo conditions for optimal cell growth as accurately as possible. For all cell-based therapeutics, i.e. a cell suspension of autologous chrondrocytes for reimplantation in a patient, the problem lies in that the end product itself cannot be sterilized, unlike certain other pharmaceutics, for instance. For this reason, guidelines such as the Good Manufacturing Practice (GMP)1, the draft guideline for Good Cell Culture Practice (GCCP)2, as well as the European Human Tissue Directive3, amongst other things, recommend the use of sterile disposable articles and/or equipment that can be sterilized for processing of human cells and tissue. Sterile conditions must be guaranteed for in vitro cell cultures throughout the entire cultivation period because, in addition to the risk of spreading contamination, the danger of infecting patients is ever present.
Clarification of Terms: Disinfection, Sterilization, DecontaminationLet us briefly look at the concepts of sterilization, decontamination and disinfection. Sterilization stands for the complete elimination and/or absence of viable microorganisms; disinfection is understood as the elimination or inactivation of all pathogens present, which, however, frequently represent only a partial quantity of all the present contaminants. The term decontamination, on the other hand, can be used in various ways, such as for the removal of biological, chemical or radioactive contamination, but it often does not allow any precise quantifiable conclusions to be made with respect to its effectiveness.
Concerning the mechanisms and verification of the effectiveness of disinfection and sterilization methods, a multitude of different guidelines and standards exists, particularly for use in the pharmaceutical industry and in the clinical sector. The pharmacopeias basically specify autoclave sterilization, hot air sterilization, and the use of ethylene oxide and sterile filtration as sterilization methods. The suitability of a specific method for a specific application must therefore be carefully scrutinized, and the sterilization process used requires validation with defined test organisms.
For effective sterilization, the various national pharmacopeias4 have agreed upon using a 6-log reduction of viable microorganisms, which is equivalent to one viable microorganism in a million, i.e. 1:1,000,000 units. This corresponds to a reduction of 99.9999% min. in the number of test organisms that were initially used.
The Development of Decontamination Concepts for CO2 IncubatorsThe various manufacturers of CO2 incubators have developed some very different concepts for the prevention and control of contamination; recently, the focus has increasingly been on process safety, effectiveness and cost awareness.
In this context, the requirement for sterility of a cell culture inside a CO2 incubator has posed significant technical challenges.
In selecting a suitable decontamination method, the following critical aspects must be taken into account:
The inner chamber of the incubator is suitable for periodic spray/wipe disinfection, which is the standard process to reduce the bio burden. Easy-to-clean metal and glass surfaces have no welding seams and, wherever possible, should have no screw connections and/or elements, which must be dismantled before disinfection to allow prompt cleaning and uniform wetting of all interior surfaces with disinfectant.
Condensation prevention, which could serve as breeding ground for germs in the incubator interior.
Secure elimination of potential contamination by means of verifiable, effective sterilization processes.
In addition, the cell culture system used should prevent the introduction of airborne germs, some of which are present even under cleanroom conditions. Cell culture bottles with a 0.2-µm bacterial filter were found to be suitable for this purpose.
The following decontamination processes can presently be found on the market:
Hot air disinfection at temperatures between 120 C and 140 C, used at different contact and cycle times (sometimes combined with HEPA filter systems), which do not represent hot air sterilization in accordance with the pharmacopeias (see Table 1). For a process using dry heat at 140 C, a 6-log reduction was indicated for B. subtilis var. Niger spores ATCC #93725.
Disinfection with wet steam at 90 C, which has shown that more thermally resistant spores may not be safely eliminated6.
A combination of wet steam 95 C/145 C hot air decontamination procedures in combination with HEPA filters, for which no studies regarding its effectiveness are available and in which filters must be regularly replaced after the decontamination procedure.
HEPA filter systems with different pore sizes, i.e. 0.3 µm, which achieve particle reduction within the incubator atmosphere7 but which also need regular maintenance.
Inner chambers made of copper to release bactericidal copper ions through oxidation, which act as cytotoxins on the respiratory chain of bacterial metabolisms. This effect has been known for hundreds of years and has been scientifically substantiated. However, this method is not suitable for all types of bacteria species, or bacteria and fungus spores, and also not for viruses. Thus it offers only limited protection. Moreover, the released copper ions are also toxic to humans.
UV treatment by application of non-ozonogenic UVC radiation with a wavelength of 253.7 nm. The mutagenic effect of UV radiation has been proven; its effectiveness, however, depends directly on direct irradiation because it has only limited penetration and thus is suitable only for the treatment of surfaces. The use of UV radiation sources for water disinfection is particularly well known. The effectiveness of treating water in humidification systems in CO2 incubators has been described8; however, it appears that additional UV treatment is not necessary if the water in the water pan is regularly replaced with sterile, distilled water (manufacturer’s recommendation is generally once to twice weekly).
Hot-air sterilization at temperatures 160 C, i.e. with dry heat, at the exposure times defined in the pharmacopeias (see Table 1). Evidence of successful sterilization of test germs pursuant to USP has been proven for single hot air sterilization programs6.
A basic requirement is that the contact period for the goods to be treated, i.e., the inner surfaces in the case of CO2 incubators, is appropriate. The cycle time can be calculated by allowing the additional times needed to heat and cool down the system.
As a rule of thumb, the higher the sterilization temperature applied, the shorter the sterilization period required.
In current publications, the focus has been on the total time of the decontamination procedure and the need for continuous contamination; any critical review should also consider the actual process time required as well as the costs for retrofitting high maintenance components, such as replacement of HEPA filters and UV lamps, as well as any costs that are associated with the recommended subsequent, manual spray/wipe disinfection. At this point, special attention should be given to the fact that the above mentioned processes, except for the hot air sterilization at temperatures of 160 C with the necessary exposure time, do not constitute processes that comply with the standards in terms of the pharmacopeias mentioned and are therefore not regarded as approved sterilization methods.
Simplicity in Application of Different Decontamination Concepts, Such As Process Safety, Effectiveness and Cost Awareness?In the following, the suitability of these processes for standards-compliant sterilization and their process safety, effectiveness, and cost awareness will be compared.
Disinfection through use of wet steam as 90 C wet heat. In this case, the advantage is that a large volume of germicidal steam can be generated with a relatively small amount of water; however, this process is not comparable to the effectiveness of autoclave sterilization with live steam at 121 C. It has been proven that the effectiveness on temperature resistant spores of species Bacillus subtilis and Bacillus stearothermophilus is unsatisfactory6,9.
At times, this required a cycle time of at least 25 hours, followed by subsequent recalibration of the CO2 sensor system. The condensate produced by the cooling of the wet steam involves potential risk of recontamination of the treated stainless steel inner surfaces. The manufacturer recommends subsequent spray/wipe disinfection with a suitable disinfectant, including the use of sterile cloths.
The use of HEPA filters to reduce particle concentration in cleanrooms and clean benches is a recognized and verifiably effective process. However, the application of HEPA filters to control contamination in CO2 incubators is contingent upon the following conditions:
- The incubator air is sucked into a HEPA filter with defined removal efficiency through suction that is generated with the fan impeller. With suitable filters, a reduction in the particle load of the incubator air can be achieved.
- In some CO2 incubators, this forced air recirculation is also required for homogenous temperature distribution and/or homogenization of the CO2 concentration within the incubator.
- However, HEPA filters still contain viable germs, which require regular filter replacement. Extended autoclaving is recommended prior to disposing the filters.
- Cycles with dry or wet steam disinfection run in CO2 incubators with HEPA filter require routine filter replacement, i.e. filter elements should be replaced prior to starting the decontamination procedure. If this decontamination procedure is carried out regularly, this can amount to considerable costs.
- When open culture flasks are handled, i.e. when moving between culture flasks, working in a low particle environment such as a clean bench is imperative to keep germs out of cultures.
- While a reduction of particles within an incubator atmosphere can minimize the contamination risk in open culture systems, such as when working with Petri dishes, it should be noted that high-quality cell culture containers with integrated bacterial filters in the screw cap are available in order to block the entry of germs from the incubator atmosphere and to prevent securely any cross contamination between the individual cell culture containers.
- Sterile conditions are absolutely essential inside cell culture containers. In this case, a reduction of particles in the atmosphere of a CO2 incubator thus would not be required.
The application of UV irradiation in combination with a copper/stainless steel alloy was described previously8. For intensive treatment of all interior surfaces, the manufacturer recommends 24-hour direct UV irradiation, if necessary, which requires prior dismantling of the sliding rack system along with the air plenum components, including the fan impeller. At the same time, all interior fittings must also be autoclaved. Following the UV cycle, all interior surfaces must once again be disinfected with 70% isopropyl alcohol and a sterile cloth. For routine application, which might be necessary at any time, this process seems relatively expensive and labor-intensive, compared to hot air sterilization using an overnight cycle. In spite of the relatively long service life of UV lighting systems (1000 hours, according to manufacturer’s specifications), the replacement of lamps would be a rather expensive item in regular application.
Minimizing Surface Contamination and Effectively Eliminating Contamination BINDER series CB and C 150 CO2 incubators are designed for easy spray/wipe disinfection and routine auto-sterilization. This customized design facilitates application and does not require replacement of expensive parts, such as filters, UV lamps, etc. It contains the following components:
Easy-to-clean, seamless, deep-drawn inner chamber with 27% less surface for potential contamination and an integrated shelf mounting system to minimize surface contamination.
Absence of condensation, even when working under conditions of highly saturated air humidity and mechanically polished stainless steel surfaces without welding seams to prevent nesting of airborne germs.
Verifiable, effective automatic hot air sterilization at 180 C in compliance with standards, which can be performed conveniently overnight and meets international guideline requirements for hot air sterilization.
References
1. http://www.fda.gov/cdrh/comp/gmp.html
2. S. Coecke et. al. Guidance on Good Cell Culture Practice, A Report of the Second ECVAM Task Force on Good Cell Culture (GCCP), ATLA 33, 261-287, 2005.
3. European Human Tissue Directive, 2006/17/EC implementing Directive 2004/23/EC of the European Parliament and of the Council as regards certain technical requirements for the donation, procurement and testing of human tissue and cells.
4. K.H. Wallhäußer, Practice of Sterilization, disinfection - Conservation, Germ Identification,
5. edition, 1995.5. J.Dalamasso, APEX Laboratories, Effective Heat Sterilization in CO2 incubators , Vol. 4, No. 3, Thermo Electron Corporation’s Heat Sterilization White Paper, 2003.
6. P. Distler, 180 C Hot air sterilization: a safe method against microbiological contamination in CO2-incubators Lab Asia, November 2003, p. 11.
7. A. Campbell, D. Figel, Importance of Class 100 Air in a CO2 incubator, Vol. 4, No. 1, Thermo Electron Corporation’s Class 100 Air White Paper, 2003.
8. H. Basujima, D. Mistry, Technical Development Report, Sanyo Electric Biomedical Co., Ltd. A Comparative Analysis of Ultra-violet Light Decontamination versus High-Heat Sterilization in the Cell Culture CO2 incubator, with the Use of Copper-Enriched Steel Construction to Achieve Background Contamination Control™, 2007.
9. Biosafety Investigation Unit, CAMR, Efficacy of a CO2 incubator heat disinfection cycle on dried microbes, 1998.
For more information, contact Daniela Maurer, Dipl. Biochem., scientific product management, BINDER GmbH, at marketing@binder-world.com or by phone at +49 7462/2005-677